                   2.32  Input # Not Assigned
                   Chapter 2: Troubleshooting Procedures

                   2.32 Input # Not Assigned


                                 1. Input # Not Assigned, displays on the Master Alarm.
                                     Yes   No
                                      ↓     → Go to “Final Actions” on page 2-7.


                                 2. The input number shown on the display is configured.
                                     Yes   No
                                      ↓     → Configure the input. See “Configuring the Master Alarm by
                                                Using the Front Panel” on page 1-63. If this solves the problem,
                                                go to “Final Actions” on page 2-7. Otherwise, call Hill-Rom
                                                Technical Support at (800) 445-3720.

                                 3. Call Hill-Rom Technical Support at (800) 445-3720.

                   2.33 Master Alarm Cannot be Browsed


                                 1. The Master Alarm cannot be browsed.
                                     Yes   No
                                      ↓     → Go to “Final Actions” on page 2-7.

                                 2. Press the Test key on the front panel. The address that shows on the
                                     display is a series: 169.254. ##. ##.
                                     Yes   No
                                      ↓     → Ensure that the Ethernet connections are properly connected on
                                                all of the alarm devices (refer to procedure 4.25 on page 4-68).
                                                They are properly connected. If this solves the problem, go to
                                                “Final Actions” on page 2-7. Otherwise, go to step 3.

                                 3. The alarm is not finding a server. Type the word Master followed by its
                                     serial number into your Internet browser’s address line: for example
                                     Master47856. You still cannot browse the Master Alarm.
                                     Yes   No
                                      ↓     → Go to “Final Actions” on page 2-7.

                                 4. Check the LED that indicates whether the alarms are communicating
                                     properly on the Master Alarm Module. See “Master Alarm Power-Up” on
                                     page 1-60. The LED is flashing at a steady rate.
                                     Yes   No
                                      ↓     → Call Hill-Rom Technical Support at (800) 445-3720.

                                 5. If you still cannot browse the Master Alarm, call Hill-Rom Technical
                                     Support at (800) 445-3720.






                   Page 2 - 38           MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)