2.27 No Display On One Area Alarm Module
Chapter 2: Troubleshooting Procedures
2.27 No Display On One Area Alarm Module
1. The displays on the other Area Alarm Modules are ON.
Yes No
↓ → Go to RAP 2.24 on page 2-30. If this solves the problem, go to
“Final Actions” on page 2-7. Otherwise, go to step 2.
2. The ribbon cable from the Area Alarm Connector P.C. Board to the Area
Alarm Module is connected.
Yes No
↓ → Connect the ribbon cable from the Area Alarm Connector P.C.
Board to the Area Alarm Module (refer to procedure 4.9 on page
4-26). If this solves the problem, go to “Final Actions” on page
2-7. Otherwise, go to step 3.
3. Replace the ribbon cable (refer to procedure 4.19 on page 4-48). The
problem is solved.
Yes No
↓ → Call Hill-Rom Technical Support at (800) 445-3720.
4. Go to “Final Actions” on page 2-7.
2.28 Audible Signal Does not Silence
1. Pressing the Silence key cancels the audible signal.
Yes No
↓ → Replace either the Master Alarm Module (refer to procedure 4.2
on page 4-10) or the Area Alarm Module (refer to procedure 4.9
on page 4-26) on the alarm panel that will not silence. If this
solves the problem, go to “Final Actions” on page 2-7.
Otherwise, call Hill-Rom Technical Support at (800) 445-3720.
2. Go to “Final Actions” on page 2-7.