                                                                     2.26  No LEDs are Lit on the Local Alarm
                                                                      Chapter 2: Troubleshooting Procedures

                              6. The wiring on the transformer is correct.

                                 Yes    No
                                  ↓      → Wire the transformer correctly. For the wiring diagram, refer to
                                            FO 3-1 at the rear of this document. If this solves the problem,
                                            go to “Final Actions” on page 2-7. Otherwise, go to step 7.
                                                                                                                 2
                              7. The voltage across pin 1 to pin 3 of TB-5 on the Local Alarm P.C. board is
                                 approximately 20 V AC.
                                 Yes    No
                                  ↓      → Replace the transformer (refer to procedure 4.13 on page 4-36).
                                            If this solves the problem, go to “Final Actions” on page 2-7.
                                            Otherwise, call Hill-Rom Technical Support at (800) 445-3720.

                              8. Go to “Final Actions” on page 2-7.