2.26 No LEDs are Lit on the Local Alarm
Chapter 2: Troubleshooting Procedures
6. The wiring on the transformer is correct.
Yes No
↓ → Wire the transformer correctly. For the wiring diagram, refer to
FO 3-1 at the rear of this document. If this solves the problem,
go to “Final Actions” on page 2-7. Otherwise, go to step 7.
2
7. The voltage across pin 1 to pin 3 of TB-5 on the Local Alarm P.C. board is
approximately 20 V AC.
Yes No
↓ → Replace the transformer (refer to procedure 4.13 on page 4-36).
If this solves the problem, go to “Final Actions” on page 2-7.
Otherwise, call Hill-Rom Technical Support at (800) 445-3720.
8. Go to “Final Actions” on page 2-7.
MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra) Page 2 - 35