                                                                                  2.34  Intermittent Browsing
                                                                      Chapter 2: Troubleshooting Procedures

               2.34 Intermittent Browsing


                              1. The Master Alarm can only be browsed intermittently.
                                 Yes    No
                                  ↓      → Go to “Final Actions” on page 2-7.
                                                                                                                 2
                              2. Ensure that the Ethernet connections are properly connected on all of the
                                 alarm devices (refer to procedure 4.25 on page 4-68). They are properly
                                 connected.

                                 Yes    No
                                  ↓      → Connect the Ethernet to all of the Alarm devices. If this solves
                                            the problem, go to “Final Actions” on page 2-7. Otherwise, go to
                                            step 3.


                              3. Check the LEDs on the Master Alarm Modules and the Area Alarm
                                 Network Communications Boards. See “Master Alarm Power-Up” on page
                                 1-60. Also see “Area Alarm Power-Up” on page 1-112. The LEDs that
                                 indicate whether the alarms are communicating are flashing very rapidly.
                                 Yes    No
                                  ↓      → Call your facility’s Information Systems personnel to fix the
                                            problem. If they fix the problem, go to “Final Actions” on page
                                            2-7. Otherwise, call Hill-Rom Technical Support at
                                            (800) 445-3720.

                              4. Have your facility’s Information Systems personnel check the ARP traffic
                                 of the facility’s network. Heavy broadcast traffic can interfere with timely
                                 Web service and data exchange between the Master Alarms and Area
                                 Alarm Network Communications Boards. The broadcast traffic is heavy.

                                 Yes    No
                                  ↓      → Call Hill-Rom Technical Support at (800) 445-3720, then go to
                                            “Final Actions” on page 2-7.

                              5. If there is excessive broadcast traffic, it usually means that there is a piece
                                 of equipment on the facility’s network that is not operating properly. The
                                 facility’s Information Systems personnel will have to eliminate some of the
                                 traffic flow, by fixing their other equipment. The browsing works correctly
                                 after the Information Systems personnel fix their other equipment.
                                 Yes    No
                                  ↓      → Call Hill-Rom Technical Support at (800) 445-3720.

                              6. Go to “Final Actions” on page 2-7.


                              NOTE:
                              ARPs are broadcasts and can consume network bandwidth.





               MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)     Page 2 - 39