2.23 Data Link Lost With Device: Device Name
Chapter 2: Troubleshooting Procedures
2.23 Data Link Lost With Device: Device Name
1. Data link lost with device: followed by the device name, displays on the
Master Alarm.
Yes No
↓ → Go to “Final Actions” on page 2-7.
2
2. The Ethernet connections are connected properly.
Yes No
↓ → Ensure all of the Ethernet connections are connected properly.
Refer to the facility’s wiring diagram of the Ethernet network. If
this solves the problem, go to “Final Actions” on page 2-7.
Otherwise, go to step 3.
3. The device is powered on.
Yes No
↓ → Turn the power to the device on. If this solves the problem, go to
“Final Actions” on page 2-7. Otherwise, go to step 4.
4. Call Hill-Rom Technical Support at (800) 445-3720.