2.4 Transducer Status Fault
Chapter 2: Troubleshooting Procedures
2.4 Transducer Status Fault
1. An Area Alarm Module displays error code A 06.
Yes No
↓ → Go to “Final Actions” on page 2-7.
2
2. Transducer is connected properly.
Yes No
↓ → Connect the transducer properly (refer to procedure 4.23 on page
4-62). If you need more information about how to connect a
transducer, refer to the MedPlus TotalAlert™ Alarm Network—
Conversion Kits Installation Instructions (is554). If this solves
the problem, go to “Final Actions” on page 2-7. Otherwise, go to
step 3.
3. Transducer has a slow steady flash on its heartbeat LED.
Yes No
↓ → Replace the transducer (refer to procedure 4.23 on page 4-62). If
this solves the problem, go to “Final Actions” on page 2-7.
Otherwise, go to step 4.
4. If the problem still exists, call Hill-Rom Technical Support at
(800) 445-3720.
MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra) Page 2 - 15