LIST OF DISCREPANCIES                               2017                               Page 1 of 2

        Facility         Florida Hospital Orlando Med Plaza  Inspector   Anthony Voight

       City/State        Orlando, FL                       Facility Contact    Darrell Huckaby
       Contractor        n/a                               Floor/Bldg    All


                 Area Served                Bed/Room #                  Discrepancy                 Disc Repair Status
       Master Alarms
       1   Area Maint. Shop Basement          Loc. #2      N2 Bank #1 In Reserve, No Visual
       2   Area Maint. Shop Basement          Loc. #2      N2 Bank #2 In Reserve, No Visual
       3   Area Maint. Shop Basement          Loc. #2      CO2 Reserve In Use, No Visual
       1  Floor
        st
       4   1.5T MRI                            ZVB         Gauges fail, Med Air reading high.
       5   Decontamination                     ZVB         No ZVB for N2 in Decontamination Room

       6   Decontamination                  Decon. Rm.     Air needs to be blanked off
                                                           Outlets have no pressure, they need to be
       7   Gamma Knife                                     blanked off if not in use.
                                                           Found light discoloration in N2O during
       8   LITHO 3                            LITHO 3      Purity Test.
                                                           Found heavy discoloration in N2O during
       9   OR 1                                OR 1        Purity Test
       10  OR 4                                OR 4        VAC outlet #3, Adapter Does Not Latch
                                                           Labeling fails, tags on piping in ZVB should
       11  PACU Beds 1-5                       ZVB         read PACU Beds 1-5
                                                           Labeling fails, tags on pipe in ZVB should
       12  PEDS 1-8                            ZVB         read Patient Rooms 1-8
           PEDS 9-12                                       Labeling fails, tags inside ZVB should read
       13  & PACU 1-2                          ZVB         PEDS 9-12 & PACU 1-2
                                                           Found light particulates/discoloration in O2
       14  Recovery Beds 6-9                   Bed 7       and Air during Purity Test
       15  Recovery Beds 6-9                   Alarm       Labeling fails.
       16  Recovery Beds 6-9                   Alarm       Normal light is out for Oxy
                                                           Found light particulates/discoloration in Air
       17  Sedation 1                        Sedation 1    during Purity Test
           Treatment Units 1 thru 4,
           Oncology Patient Hold, Exam                     Labeling fails, should read Treatment Units
           Rooms 1 thru 6, 8 & 9 Virtual                   1 thru 4, Oncology Patient Hold, Exam
           Simulation, Meg Suite, & Soiled                 Rooms 1 thru 6, 8 & 9 Virtual Simulation,
       18  Utility                           ZVB/Alarm     Meg Suite, & Soiled Utility
       19  X-ray                               X-ray       VAC outlet #1, Adapter Does Not Latch