LIST OF DISCREPANCIES                               2017                               Page 2 of 2


                  Area Served                Bed/Room #                 Discrepancy                 Disc Repair Status
       3  Floor Continued
        rd
           Consult Rms 2-4, Proc. Rms. 1-2
       20  and Holding Ste 331                  ZVB        Label needs to include Holding
           Consult Rms 2-4, Procedure                      Labeling fails, it needs to drop X-ray 3-4
       21  Rooms 1-2, Holding (Suite 331)   ZVB (Air/VAC)  and CT Scan
           Consult Rms 2-4, Procedure                      Labeling fails, ZVB is not labeled for room
       22  Rooms 1-2, Holding (Suite 331)     ZVB (N2)     served
           Consult Rms 2-4, Procedure
       23  Rooms 1-2, Holding (Suite 331)     ZVB (N2)     Gauges fail, N2 reading high
           Consult Rms 2-4, Procedure
       24  Rooms 1-2, Holding (Suite 331)      X-ray 1     VAC outlet #1, Leaks WithOut Adapter In
           Consult Rms 2-4, Procedure
       25  Rooms 1-2, Holding (Suite 331)     Radiology    VAC outlet #1, Adapter Does Not Latch
        th
       4  Floor
       26  OR 1                                 OR 1       VAC outlet #2, Leaks WithOut Adapter In
       27  OR 4                                 ZVB        Leak Test fails, gauges are leaking
       28  OR 4                                 OR 4       VAC outlet #2, Leaks WithOut Adapter In
                                                           WAGD outlet #1, Leaks WithOut Adapter
       29  OR 4                                 OR 4       In
       30  OR 5                                 ZVB        Missing N2 tag on the outside
                                                           VAC outlet #2, has Low Flow and Leaks
       31  OR 5                                 OR 5       WithOut Adapter In
       32  OR 7                                 ZVB        N2 is missing label
                                                           VAC outlet #1 has Low Flow and Leaks
       33  OR 7                                 OR 7       WithOut Adapter In
       34  OR 7                                 OR 7       VAC outlet #2, Leaks WithOut Adapter In
       35  Stage 1 Recovery                     Alarm      Oxy fails, alarmed at 32

       36  Stage 1 Recovery                     Alarm      Air fails, alarmed at 30
        th
       5  Floor
           Children’s Center for Cancer and                Labeling fails, should read Children’s
           Blood Diseases (Ste. 589) Exam                  Center for Cancer and Blood Diseases
           Rms 1-3, 7-8, Pediatric Cellular                (Ste. 589) Exam Rms 1-3, 7-8, Pediatric
           Therapy (Ste. 581) Exam Rms 1-                  Cellular Therapy (Ste. 581) Exam Rms 1-
           5, Adler Family Coordinated Care                5, Adler Family Coordinated Care for Kids
       37  for Kids (Ste. 586) Room 3           ZVB        (Ste. 586) Room 3
        th
       6  Floor
           Suite 683 Cancer Institute of
       38  Florida Exam 1-4                     ZVB        There is no ZVB for this area