4.25 Internet Connections
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.25 Internet Connections
Tools required: 5/32" hex wrench
Removal
1. Using a 5/32" hex wrench, unlock the front panel (A) (see figure 4-26 on
page 4-68).
Figure 4-26. Internet Connections
SHOCK HAZARD:
Turn off the power supply in the alarm, and leave it off during the
installation of any component or sub-assembly. Failure to do so could
result in personal injury or equipment damage.
2. Turn the power switch (D) to the OFF position.
3. Depending on the alarm, unplug the Ethernet cable from the Master Alarm
Module (B) or from the Area Alarm Network Communications Board (C).
Page 4 - 68 MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)