                   4.25  Internet Connections
                   Chapter 4: Removal, Replacement, and Adjustment Procedures

                   4.25 Internet Connections

                                 Tools required:   5/32" hex wrench


                          Removal

                                 1. Using a 5/32" hex wrench, unlock the front panel (A) (see figure 4-26 on
                                     page 4-68).

                                               Figure 4-26. Internet Connections









































                                 SHOCK HAZARD:
                                 Turn off the power supply in the alarm, and leave it off during the
                                 installation of any component or sub-assembly. Failure to do so could
                                 result in personal injury or equipment damage.

                                 2. Turn the power switch (D) to the OFF position.


                                 3. Depending on the alarm, unplug the Ethernet cable from the Master Alarm
                                     Module (B) or from the Area Alarm Network Communications Board (C).









                   Page 4 - 68           MedPlus TotalAlert™ Alarm Network Operation and Maintenance Manual (man291ra)