“Oil-Less” Scroll Medical Air


            6.0  Inspection/Replacement Procedures



            6.2  Air Intake Filter                                               Canister Housing


                               WARNING:                               Protective Cover
               Before    starting    any    maintenance
               procedures, disconnect all power to the
               package.




            Release all pressure from the package
            before removing, loosening, or servicing any
            covers, guards, fi ttings, connections, or other
            devices.


            Never perform any maintenance functions
            while the unit is in operation.                                              Element
                                                                 Figure 6.3 Air Intake Filter
            The air intake fi lter element should be changed
            annually under normal operating conditions.  To
            change the fi lter:

            1. Turn  off  the compressor being serviced and
               lock open the appropriate disconnect switches.
            2.  Close intake isolation valve.
            3.  Remove the protective cover by loosening the
               wing nut (if applicable) and latches.

            4.  Remove the element.
            5.  Clean inside of housing
            6.  Insert a new element (note orientation of the
               element).
            7.  Replace protective cover and tighten wing nut
               (if applicable) and latches.
            8.  Open intake isolation valve.

            9.  Turn on the compressor.

















                                                             6-3                                  4107 9000 69.06