“Oil-Less” Scroll Medical Air
6.0 Inspection/Replacement Procedures
6.2 Air Intake Filter Canister Housing
WARNING: Protective Cover
Before starting any maintenance
procedures, disconnect all power to the
package.
Release all pressure from the package
before removing, loosening, or servicing any
covers, guards, fi ttings, connections, or other
devices.
Never perform any maintenance functions
while the unit is in operation. Element
Figure 6.3 Air Intake Filter
The air intake fi lter element should be changed
annually under normal operating conditions. To
change the fi lter:
1. Turn off the compressor being serviced and
lock open the appropriate disconnect switches.
2. Close intake isolation valve.
3. Remove the protective cover by loosening the
wing nut (if applicable) and latches.
4. Remove the element.
5. Clean inside of housing
6. Insert a new element (note orientation of the
element).
7. Replace protective cover and tighten wing nut
(if applicable) and latches.
8. Open intake isolation valve.
9. Turn on the compressor.