Page 209 - Womens Pavilion
P. 209

“Oil-Less” Scroll Medical Air


           6.0  Inspection/Replacement Procedures



           6.6  Backup Pressure Switch Set Point               Adjusting Instructions
           Adjustments
                                                               1.  To locate the adjuster, slide the adjustment
                                                                   access cover to reveal a slotted adjustment
           The backup switch is set at the factory to the
           operating point(s) as stated on the wiring diagram      screw.
           supplied with the unit.  It is good practice to cycle   2.  Turn the screw inward (clockwise) to increase
           the switch to determine  actual operating points
           before proceeding with readjustment.                    the setpoint and outward (counter-clockwise)
                                                                   to decrease the setpoint.  The backup pressure
                                                                   switch should always be set with falling
           CAUTION:
                                                                   pressure level starting at a pressure level higher
                                                                   than the setpoint.
           •  ALWAYS change pressure setting gradually.
           •  ALWAYS check switch setting before making        3.  Using the pressure gage, determine the
              any adjustments.                                     actuation point of the switch.

           •  DO NOT force slotted adjustment screw when       4.  If the actuation point is above the desired value,
              it becomes diffi  cult to turn.                      turn the slotted adjustment screw counter-
                                                                   clockwise to decrease the actuation point, and
           •  ALWAYS isolate the pressure transducer               if it is below, turn the slotted adjustment screw
              before making any adjustments to the backup          clockwise to increase it.
              pressure switch.
                                                               5.  For exact pressure setting, cycle pressure switch
                                                                   and make fi ne adjustments be repeating steps
                  Adjustment Access Cover                          2 through 4 (trial and error process) until the
                                                                   desired setting is obtained.

                                                               6.  Slide the adjustment access cover back into
                                                                   place over the slotted adjustment screw.








                              Increase












                                 Slotted Adjustment Screw
                                         (Internal)


           Figure 6.5 Backup Pressure Switch



           4107 9000 69.06                                  6-6