“Oil-Less” Scroll Medical Air


           Appendix A: Desiccant Dryer



           A.2  Operation                                      11. During the conditioning run test all joints to
                                                                   locate any leaks using leak detector spray or a
           A.2.1   Initial Start-Up                                suitable alternative. Tighten or repair any leaks
                                                                   and retest.

                                                               NOTE:  Any small leaks on the dryer outlet side
                                                               will cause a deterioration of the dew point.

                                                               12. Press “Automatic” on the dryer display screen
                                                                   to begin operation in Automatic mode.

                                                               13. On the completion of the conditioning run,
                                                                   slowly open the source isolation valve. The
                                                                   dryer will now be fully operational.

                                                               A.2.2   Procedure to Switch Off  Dryer


                                                               1.  Put second dryer on line by repeating steps 4
                                                                   thru 7 above.
           Figure A.1 Main Screen - Dryer Operation
                                                               2.  CLOSE the dryer outlet isolation valve.
           1.  Switch on the electrical supply to the dryer.

                                                               3.  Press “OFF” on the dryer display screen.
           2.  CLOSE the dryer isolation valves.
                                                               4.  CLOSE the dryer inlet isolation valve. Dryer
           3.  Check that the compressed air supply is on.         should de-pressurize.
              Let the system come up to pressure

           4.  Slowly OPEN the dryer inlet isolation valve.                       WARNING:
                                                                 Wait at least 2 minutes for pressure in
           5.  Press “Manual” on the dryer display screen to
              begin operation.  The dryer will now begin to      the dryer to decay before performing any
              cycle.                                             service to the dryer.


           6.  Check that purge air is fl owing from the purge   A.2.3   Normal Start-up
              muffl  er.
                                                               This procedure is to be followed when the dryer
           7.  Slowly OPEN the dryer outlet isolation valve.
                                                               has been shut down for a short period during
           8. Open the dew point and CO sensor (if             which time the desiccant has not been exposed
              supplied) isolation valves.                      to wet gas.

           9.  Check for airfl ow at the dew point sensor       1.  Start up the compressor if shut down.
              orifi ce.
                                                               2.  Set the appropriate dryer to Automatic mode.
           10. Operate the dryer for fi ve to ten minutes with
              the source isolation valve closed.               3.  Slowly OPEN the dryer inlet isolation valve.
                                                               4.  Slowly OPEN the dryer outlet isolation valve.


                                                               5.  Check operation of the dryer.
           4107 9000 69.06                                  A-2