Page 228 - Womens Pavilion
P. 228

“Oil-Less” Scroll Medical Air


            Appendix B: Dew Point Transmitter



            B.1  General Information                             B.2  Introduction

            CAUTION:  This manual is designed to serve as        The dew point transmitter is a continuous, on-line
            the operation and maintenance guide for your         instrument that measures the absolute moisture
            Dew Point Transmitter, if equipped. The contents     content in the fi nal air line. The transmitter
            of this manual should be carefully read BEFORE       measures dew point with excellent long term
            attempting any phase of operation or maintenance.    stability. The Advanced Ceramic Moisture sensor
            Failure to follow the operating and maintenance      is durable and has been designed for ruggedness
            procedures of the instruction manual could result    and simplicity. The transmitter is fully calibrated at
            in personal injury or property damage.               the factory prior to shipment.


                               WARNING:                          B.3  Specifi cations
             Before      starting     any     installation,      1.  Dew point Temperature: -100° to 20°C (-148°
             maintenance       or    service   procedure,           to 68°F)
             disconnect ALL power to  the system to
             prevent electrical shock.                           2.  Operating Temperature: 0° to 60°c (32° to
                                                                    140°F)

             Before making or breaking any medical               3.  Dew point accuracy: ±2°C (±3.6°F)
             gas line connections, make sure the                 4.  Air Consumption:  0.75 LPM (1.6 SCFH)
             system is depressurized in order to avoid
             personal injury.                                    B.3.1   Output


             Before removing the dew point transmitter,          Analog output: 4 - 20 mA
             verify that the source of line pressure has
             been closed and the line pressure reduced           B.3.2   General
             to atmospheric pressure.
                                                                 1.  Operation Voltage: 12 - 28 VDC
             An alarm condition on dew point                     2.  Probe material: Stainless Steel (316)
             indicates a dew point level exceeding the           3.  Sensor protection: Ceramic
             maximum set point, or a faulty dew point
             transmitter. Immediate action should be             B.4  Operation
             taken to reduce the possibility of high
             dew point in the Medical Air line.                  Although the correct operation of the transmitter
                                                                 is not sample fl ow dependent, it is important that
             If the dew point transmitter fl owmeter              fl ow velocity through the sample source to the
             becomes clogged, dew point readings                 sample block is high enough to avoid long lead
             may be inaccurate, allowing moisture to             time lags in response to changes in moisture at the
             accumulate undetected.                              sample source.












                                                             B-1                                  4107 9000 69.06