Page 233 - Womens Pavilion
P. 233
“Oil-Less” Scroll Medical Air
Appendix C: CO Transmitter
C.6 Operation To replace the sensor (see Fig C.1), disconnect
the power to the unit, disconnect the inlet air
CAUTION: At initial startup, if the unit is reading connection and unscrew the cover to access
a gas level, do not make any adjustments for a few the inside of the transmitter. Next, unplug the
hours until the unit has a chance to settle in and CO sensor from the bottom of the internal PC
stabilize. board. Then remove the cap on the bottom of the
transmitter and remove the CO sensor and wiring.
If gas readings remain high or below zero Unplug the wiring from the sensor and replace it
(-0), re-calibration may be needed. We also with a new one. Reinstall the sensor in the reverse
recommend checking the compressor’s air intake order. Once the sensor is installed go through
for contamination fi rst. steps C.9.1 and C.9.2 to calibrate and verify the
proper operation of the CO sensor.
C.7 Maintenance
C.9 Calibration
WARNING:
C.9.1 Zeroing Transmitter
Electrical Shock Hazard. No repair should
ever be attempted by anyone not having The following procedure should be used for
experience in the repair of devices of this zeroing the transmitter. (See C.11 Accessories and
nature. Failure to follow proper repair Replacement Parts for zero gas part numbers)
procedures can result in serious injury.
• Turn on the power and allow the transmitter
to warm up for several hours to stabilize.
Replace damaged parts with components from
BeaconMedæs. Test the unit after installation of • Disconnect the inlet air line to the transmitter
replacement parts to make certain that it complies and connect the air line from the zero gas
with the published specifi cations. Contact cylinder.
BeaconMedæs Technical Services department at
1-888-4MEDGAS (1-888-463-3427) for technical • Allow the calibration gas from the cylinder to
support or to order spare parts. fl ow through the sensor for approximately 3
minutes to stabilize the sensor.
C.7.1 Maintenance Schedule
• Remove the cover from the transmitter housing.
Maintenance Frequency Action
• Locate the blue potentiometer, marked “Z”
Recalibration Every 6 See Section C.9 (top potentiometer) and adjust the pot until
months the I2 reading on the main diagnostic screen
reads 700. Turning the adjusting screw
Replace Every 2 years See Section C.8 counterclockwise will reduce the reading.
Sensor Turning it clockwise will raise the reading.
• Proceed to C.9.2 for calibrating the transmitter.
C.8 Sensor Checkout & Replacement
To check a sensor’s response, test gas has to be
placed on the sensor. When it fails to show a
gas response during calibration, a new sensor is
required.
4107 9000 69.06 C-2