Page 233 - Womens Pavilion
P. 233

“Oil-Less” Scroll Medical Air


           Appendix C: CO Transmitter



           C.6  Operation                                      To replace the sensor (see Fig C.1), disconnect
                                                               the power to the unit, disconnect the inlet air
           CAUTION: At initial startup, if the unit is reading   connection and unscrew the cover to access
           a gas level, do not make any adjustments for a few   the inside of the transmitter.  Next, unplug the
           hours until the unit has a chance to settle in and   CO sensor from the bottom of the internal PC
           stabilize.                                          board. Then remove the cap on the bottom of the
                                                               transmitter and remove the CO sensor and wiring.
           If gas readings remain high or below zero           Unplug the wiring from the sensor and replace it
           (-0), re-calibration may be needed.  We also        with a new one. Reinstall the sensor in the reverse
           recommend checking the compressor’s air intake      order. Once the sensor is installed go through
           for contamination fi rst.                            steps C.9.1 and C.9.2 to calibrate and verify the
                                                               proper operation of the CO sensor.
           C.7   Maintenance
                                                               C.9  Calibration

                             WARNING:
                                                               C.9.1   Zeroing Transmitter
            Electrical Shock Hazard.  No repair should
            ever be attempted by anyone not having             The following procedure should be used for
            experience in the repair of devices of this        zeroing the transmitter. (See C.11 Accessories and
            nature.  Failure to follow proper repair           Replacement Parts for zero gas part numbers)
            procedures can result in serious injury.
                                                               •  Turn on the power and allow the transmitter
                                                                   to warm up for several hours to stabilize.
           Replace damaged parts with components from
           BeaconMedæs. Test the unit after installation of    •  Disconnect the inlet air line to the transmitter
           replacement parts to make certain that it complies      and connect the air line from the zero gas
           with the published specifi cations.  Contact             cylinder.
           BeaconMedæs Technical Services department at
           1-888-4MEDGAS (1-888-463-3427) for technical        •  Allow the calibration gas from the cylinder to
           support or to order spare parts.                        fl ow through the sensor for approximately 3
                                                                   minutes to stabilize the sensor.
           C.7.1   Maintenance Schedule
                                                               •  Remove the cover from the transmitter housing.
           Maintenance Frequency          Action
                                                               •  Locate the blue potentiometer, marked “Z”
           Recalibration   Every 6        See Section C.9          (top potentiometer) and adjust the pot until
                           months                                  the I2 reading on the main diagnostic screen
                                                                   reads 700.  Turning the adjusting screw
           Replace         Every 2 years  See Section C.8          counterclockwise will reduce the reading.
           Sensor                                                  Turning it clockwise will raise the reading.


                                                               •  Proceed to C.9.2 for calibrating the transmitter.
           C.8  Sensor Checkout & Replacement

           To check a sensor’s response, test gas has to be
           placed on the sensor.  When it fails to show a
           gas response during calibration, a new sensor is
           required.

           4107 9000 69.06                                  C-2
   228   229   230   231   232   233   234   235   236   237   238