“Oil-Less” Scroll Medical Air
Appendix D: TotalAlert Embedded Control System
D.2.2 5.7” User Interface for Source Systems
Service
Alarms (no Shutdown)
Shutdown
Event History Log
System Confi guration Settings
Figure D.3 Main Screen
Horn Silence
The primary master screen user interface is
displayed on a 5.7” 640 x 480 pixel display as
shown in Figure D.3. The interface is designed D.2.3 5.7” Boot/Communication Screen
such that any information can be accessed with a
minimal amount of touches by the user.
The 5.7” screen is divided into two main areas –
the top portion above the toolbar which changes
depending on the icon selected on the toolbar
and bottom portion which contains the toolbar
(Figure D.4) and is available on most screens.
Figure D.5 Boot/Communication Screen
Figure D.4 5.7” Screen Toolbar
The boot/communication screen (Figure D.5)
From left to right, the toolbar icons represent the
following: shows at boot time and will change to the
default main screen once communication and
Main Screen (default) ** If No Dryers compatibility are confi rmed:
(Image is 2x wide and Dryer Image
is not shown). • If a RS485 link failure is detected the link icon
at the top of the screen will appear and the
** Dryer Information (Medical Air with bottom version button will appear to allow
Dryers Only) the user to check display board (PCB1) version
information.
D-3 4107 9000 69.06