“Oil-Less” Scroll Medical Air
Appendix D: TotalAlert Embedded Control System
• If the connecting board is incompatible • If a unit is unavailable because it is in Manual
with the display board (for example, board mode and it is running, that color bar shows
not properly connected), only the bottom green and the adjoining symbol shows “X”.
version button will appear with a red button
background. • If the unit is unavailable because the unit is
an expandable unit, that color bar shows dark
D.2.4 5.7” Main Screen grey and the adjoining symbol shows “E”.
• If the unit is unavailable because the unit
cannot be reached via ethernet, that color
bar shows dark grey and the adjoining symbol
shows “?”.
D.2.5 5.7” Trend Screen
Figure D.6 Main Screen
The main (default) screen (Figure D.6) shows the
pertinent system measurements as well as unit
sequence information.
The pertinent system measurements include:
Pressure, Dew Point, and CO Level. Pressing Figure D.7 Trend Screen
pertinent system data buttons shows Trend
information for that value. The trend screen (Figure D.7) shows the measured
value over a specifi c time period:
Unit Sequence information shows all system units.
• The default time period when opening the
• If the unit is running, that color bar shows window is the last 60 minutes. For a given
green and the adjoining symbol shows “*” for period, the maximum amount of data stored
across the line (full speed). will only be for the most recent time period
and the older data will be removed from view
• If a unit is available and not running, that color and memory.
bar shows blue and the adjoining symbol
shows “+” if that unit is next to start. • Another time period is selected by pressing the
buttons at the bottom of the window. These
• If a unit is stopped, that color bar shows red
and the adjoining symbol shows “!” if stopped periods are 60 minutes (240 data points – 0.25
for a shutdown. min resolution), 4 hours (240 data points – 1
min resolution), 24 hours (240 data points – 6
• If the unit is unavailable and not running, that min resolution) and 6 days (240 data points –
color bar shows dark grey and the adjoining 0.6 hour resolution).
symbol shows “X”.
4107 9000 69.06 D-4