“Oil-Less” Scroll Medical Air
Appendix B: Dew Point Transmitter
B.5 Alarms
WARNING:
Dew points that exceed set points shall cause an Electrical shock hazard
alarm condition at the control panel. When the
dew point exceeds the alarm set point, the alarm No repair should ever be attempted by
contacts are de-energized. The alarm remains de- anyone not having experience in the
energized until the alarm condition is cleared by repair of devices of this nature. Failure
the operator. A high dewpoint alarm will activate to follow proper repair procedures can
if the transmitter loses power or is disconnected result in serious injury.
from the control system.
CAUTION: No maintenance and testing should
WARNING:
ever be undertaken or attempted by anyone not
Respond to alarm conditions immediately. having general experience in the repair of devices
An alarm condition on the dew point of this nature. Also, to avoid damaging the unit or
indicates a dew point level exceeding any of its components, no maintenance and testing
the maximum set point. Immediate should be undertaken by qualifi ed individuals
action to correct the problem should be who are not familiar with the procedures in this
taken. Prolonged exposure to condensing manual.
moisture can damage the Medical Air
equipment. Replace damaged parts with components from
BeaconMedæs. Test the unit after installation of
replacement parts to make certain that it complies
B.6 Maintenance with the published specifi cations.
B.6.1 Repair Policy B.6.2 Maintenance Schedule
Do not use a unit that is not functioning properly Maintenance Frequency Action
until all necessary repairs have been made
and the unit has been tested to determine Check fl ow Weekly Check for proper
that it is functioning in accordance with the through fl ow
manufacturer’s published specifi cations. Contact orifi ce
BeaconMedæs Technical Services department at
1-888-4MEDGAS (888-463-3427) for assistance. Check Yearly Verify dew point
transmitter sensor accuracy
NOTE: To ensure full reliability, have all accuracy (contact
maintenance and testing done by a qualifi ed BeaconMedæs)
technician. If this cannot be done, maintenance
and testing of those parts discussed in this manual Replace Every 2 years See Section B.9
may be undertaken by a competent, trained Sensor
individual having experience in the repair of
devices of this nature.