Page 252 - Womens Pavilion
P. 252

“Oil-Less” Scroll Medical Air


            Appendix D:  TotalAlert Embedded Control System





                                                                 D.7.7   Network Setup

                                                                 This Network Setup page (Figure D.33) is used
                                                                 to confi gure the facility network information and
                                                                 e-mail server confi guration.

                                                                 NOTE: Obtain the IP Address and DNS Name from
                                                                 the facility’s Information Systems department.
















            Figure D.31 Setup Menu

            D.7.6   Device Setup

            This Device Setup page (Figure D.32) is used to      Figure D.33 Network Setup
            confi gure the medical air system name, location,
            facility name, and contact information.
                                                                 1.  Click Network Setup  to access the Network
                                                                    Setup page.

                                                                 2.  Select Enable or Disable for the DHCP Mode.


                                                                 3.  Enter the IP Address for the DNS server
                                                                    confi guration.

                                                                 4.  Enter the IP Address and DNS name for the
                                                                    e-mail SMTP server confi guration.
            Figure D.32 Device Setup
                                                                 5.  Click the Submit button.
            1.  Click Device Setup to access the Device Setup
               page.                                             D.7.8   Website Setup

            2.  Enter the new device name.                       This Website Setup page (Figure D.34) is used to
                                                                 confi gure the automatic refresh interval for the
            3.  Enter the location                               website.  By selecting an auto-refresh interval, the
                                                                 web page will be refreshed automatically, keeping
            4.  Enter the facility description.
                                                                 all information current during a viewing period.
            5.  Enter the contact information.

            6.  Click the Submit button.
                                                             D-17                                 4107 9000 69.06