Page 252 - Womens Pavilion
P. 252
“Oil-Less” Scroll Medical Air
Appendix D: TotalAlert Embedded Control System
D.7.7 Network Setup
This Network Setup page (Figure D.33) is used
to confi gure the facility network information and
e-mail server confi guration.
NOTE: Obtain the IP Address and DNS Name from
the facility’s Information Systems department.
Figure D.31 Setup Menu
D.7.6 Device Setup
This Device Setup page (Figure D.32) is used to Figure D.33 Network Setup
confi gure the medical air system name, location,
facility name, and contact information.
1. Click Network Setup to access the Network
Setup page.
2. Select Enable or Disable for the DHCP Mode.
3. Enter the IP Address for the DNS server
confi guration.
4. Enter the IP Address and DNS name for the
e-mail SMTP server confi guration.
Figure D.32 Device Setup
5. Click the Submit button.
1. Click Device Setup to access the Device Setup
page. D.7.8 Website Setup
2. Enter the new device name. This Website Setup page (Figure D.34) is used to
confi gure the automatic refresh interval for the
3. Enter the location website. By selecting an auto-refresh interval, the
web page will be refreshed automatically, keeping
4. Enter the facility description.
all information current during a viewing period.
5. Enter the contact information.
6. Click the Submit button.
D-17 4107 9000 69.06