“Oil-Less” Scroll Medical Air
Appendix D: TotalAlert Embedded Control System
By selecting the activity that is overdue (in red), D.7.2 Set Up: Physical Connection
the detail screen appears for that service item and
provides the necessary information for the service WARNING:
activity: part number, description.
ONLY CONNECT THE CUSTOMER
D.6.3 Resetting a Service Activity NETWORK TO THE CUSTOMER
ETHERNET CONNECTION AS SHOWN
IN FIGURE D.27.
DO NOT UNPLUG EXISTING ETHERNET
CABLES USED FOR THE INTERNAL
CONNECTIONS. DO NOT ATTEMPT
TO REMOVE ETHERNET SECURITY
PROTECTORS TO USE FOR CUSTOMER
ETHERNET CONNECTION.
IF THE CUSTOMER ETHERNET IS
PLACED IN AN INTERNAL ETHERNET
CONNECTION, THE INTERNAL
ETHERNET FUNCTIONALITY WILL NOT
Figure D.26 Resetting a Service Activity PERFORM AS DESIGNED.
After a service activity is performed, to reset the
schedule, do the following:
• Select the service activity on the touchscreen
(Figure D.26).
• Press the Reset button.
• Enter the password (as explained in Section
D.4).
D.7 Remote Monitoring
Customer Ethernet Connection
CAUTION: The information systems personnel on PCB3 Master
should be notifi ed before changing any of the
network settings. Changing the settings could Figure D.27 Connecting the cable
keep the equipment from working properly.
1. Using a Cat5 Ethernet cable, connect the
D.7.1 Set Up: Equipment Required medical air system to an Ethernet switch or
hub. Connect the cable to the Customer
• PC with an Ethernet connection Ethernet Connection (Figure D.27) on PCB3
Master board.
• PC with a web browser, such as Microsoft
Internet Explorer 2. Verify the green LINK LED on the printed circuit
board illuminates.
• Cat5 or better Ethernet cable
4107 9000 69.06 D-14