“Oil-Less” Scroll Medical Air


           Appendix D:  TotalAlert Embedded Control System



           D.4  Password Access                                •  Maintenance - Testing the System Alarms:  The
                                                                   Testing button allows the operator to create
                                                                   an Alarm or Shutdown condition to test the
                                                                   complete alarm network.  See section D.5 for
                                                                   additional information.

                                                               •  Settings - Pressure: Adjust system “Max” and
                                                                   “Min” operating pressure settings within a
                                                                   factory-set range of pressures.

                                                               •  Settings - Horn: An option is available to
                                                                   reinitiate the horn during an alarm/shutdown
                                                                   event.  The operator can elect to have the horn
                                                                   reinitiate after a set period of time after the
                                                                   silencing of the horn.
           Figure D.24 Password Screen
                                                               •  Settings - Language: The operator can select
                                                                   from the language options available on the
           The TotalAlert Embedded control system contains
           items that can be changed or adjusted.  Before          system.
           making any changes on the system or unit screens,   •  Settings - Display Units:  The operator can
           a password prompt appears.  A proper password
           must be entered before adjusting the control            select preferences for the display of pressure,
           system.  See Figure D.24.                               dew point, and other temperatures.

                                                               •  Settings - Date and Time:  Ensure the correct
           The password for the Scroll Medical Air System
           is “121212”.   After entering the password, the         date and time for the system, as this impacts
                                                                   history log reporting.
           password remains valid for approximately ten
           minutes or until the unit resets itself to the default   •  Settings - Network Confi guration:  Confi gure
           display screens.  Reentering the password after         how the system operates remotely within the
           “timing out” allows the user to continue making         existing facility network.
           system adjustments.
                                                               •  Settings - General Options: Select method for
           D.4.1   Main Screen Password Items
                                                                   displaying alarms/shutdowns.  Operator can
                                                                   change the timer setting for sleep mode on the
           On the 5.7” main screen, several screens allow
           adjustments with the proper password.                   screens.
                                                               D.4.2   Unit Screen Password Items
           •  Maintenance - Resetting a Service Activity:  A
              service notifi cation activates when a service    In the Service section of the Unit screens, the
              item is past due.  Upon completion of the        operator can create alarm or shutdown conditions
              past due service activity, reset the service due   to test the complete alarm network.  See section
              counter by pressing the item on the service      D.5 for additional information.
              screen, selecting update, and entering the
              password when prompted.







           4107 9000 69.06                                 D-12