Page 292 - Womens Pavilion
P. 292

Oil-Less Claw Medical Vacuum


            6.0  Maintenance


            6.1.3   Gear Box Lubrication                         Standard Claw
                                                                 The back side bearings of the 15 HP vacuum units
            CAUTION: Oil and used oil must be disposed of        need to be greased every 6 months.  Use a grease
            corresponding with the relevant health, safety and   gun to fi ll until grease fl ows from relief port (see
            environmental laws.                                  2 greasing points (7) (Ref. Fig 6.1).  The greasing
                                                                 interval should be cut in half when operating at
            Check the oil level regularly by observing the       high ambient temperatures.  See Section 7.1 for
            built-in sight glass (4).  Each pump requires the    listing of grease service kits for Standard Claw
            proper oil to operate correctly (see Section 6.1.3   Systems.
            for the correct oil types).  The oil should be
            changed every 10,000 hours of operation.  The        O Assured Claw
                                                                  2
            oil will need to be changed more frequently          The back side bearings of the 15 HP vacuum units
            when operating at ambient temperatures above         need to be greased every 6 months.  Each bearing
            105°F.  An oil analysis kit is available to check oil   requires 50 grams of grease.
            quality and help determine best service interval
            for the operating conditions.  See Oil Analysis Kit   Remove fan cover on the end of the pump opposite
            in Sections 7.1 and 7.2.  Contact BeaconMedæs        the motor.  Follow this procedure for each bearing:
            Tech Support for additional information.
                                                                 1.  Remove the bearing cap.
            The following table describes the oil quantity       2.  Clean all old grease from the cap and face of
            capacity per pump.  See the pump nameplate for          the bearing.
            the pump model number.
                                                                 3.  Use a clean cloth and wear latex gloves, so
             Table 6.2 Oil Quantity per Pump                        that no contaminates are introduced into the
                                                                    bearing.
                                  Oil Quantity (liters)
                                                                 4.  Pack 6.5 grams of grease into the bearing (using
              Pump Model         Standard     O Assured
                                                2                   your fi nger).
                 MM1104             .85           .85
                                                                 5.  Pack the balance of the grease (43.5 grams)
                 MM1144             .85           .85
                                                                    into the cap.
                 MM1102             .85           .85
                                                                 6.  Reinstall the bearing cap.
                 MM1142             .85           .85
                 MM1202             1.0           1.0            7.  Reinstall the fan cover.
                 MM1252             1.0           1.0            The greasing interval should be cut in half when
                 MM1322             1.0           1.0            operating at high ambient temperatures.
                 MM1402             1.0           1.0
                                                                 Use only BeaconMedæs O Assured grease.  Failure
                                                                                          2
                 MI1502             .6             .6            to use the proper grease may void the warranty.
                                                                 BeaconMedæs O Assured grease is inert in the
                                                                                   2
            6.1.4   Bearing Lubrication                          presence of oxygen that may be in your vacuum
                                                                 system.
            CAUTION:  Shut down the vacuum pump and
            lock & tag out the electrical controls for that pump   See Section 7.1 for listing of grease service kits for
            prior to greasing the pump bearings.                 O Assured Claw Systems.
                                                                  2






                                                             6-3                                  4107 9000 95.04
   287   288   289   290   291   292   293   294   295   296   297