Page 295 - Womens Pavilion
P. 295
Oil-Less Claw Medical Vacuum
7.0 Replacement / Maintenance Parts
7.2 Service Kits for O2Assured Claw Systems
KIT NUMBER DESCRIPTION QTY WHERE USED CONTENTS
6-Month Basic Service Kit for 15 Hp O2Assured Claw Systems
KIT - Claw Vacuum Pump 1 per (1) High melting point
4107 4001 25 15 Hp B-Side Bearings
Bearing Grease O2Assured pump grease (400g tube)
1-Year Basic Service Kit for O2Assured Claw Systems
KIT - Claw Vacuum Basic 1 per
4107 4001 19 2 - 5.4 Hp Filter element
Size A O2Assured pump
KIT - Claw Vacuum Basic 1 per
4107 4001 20 6.4 - 15 Hp Filter element
Size B O2Assured pump
10,000 Hour Oil Change Kits for O2Assured Claw Systems
(1) Vent Valve
KIT - Claw Vacuum Oil 1 per
4107 4001 21 2 - 5.4 Hp (2 Pt) Synchronising
Change Size A O2Assured pump
Gear Operating Fluid
(1) Vent Valve
KIT - Claw Vacuum Oil 1 per
4107 4001 22 6.4 - 15 Hp (2 Pt) Synchronising
Change Size B O2Assured pump
Gear Operating Fluid
Oil Analysis Kit for O2 Assured Claw Systems
KIT - Claw Vacuum Oil 1 per (1) Oil sample container
4107 4001 27 All systems
Analysis O2 Assured pump (1) Instruction form
Replacement Parts
1 per
4107 6541 85 Coupling Insert 2 Hp - 7.5 Hp (1) Coupling insert
pump
1 per 8.7 Hp - 10* Hp
4107 6541 86 Coupling Insert (1) Coupling insert
pump (MM1322)
1 per
4107 6541 87 Coupling Insert 10* Hp (MM1402) (1) Coupling insert
pump
1 per
4107 6541 88 Coupling (Complete) 15 Hp (1) Complete coupling
pump
Note: There are two models of 10 Hp Claw pumps: MM1322 and MM1402. See pump nameplate to
identify pump model.
4107 9000 95.04 7-2