Page 295 - Womens Pavilion
P. 295

Oil-Less Claw Medical Vacuum


           7.0  Replacement / Maintenance Parts


           7.2  Service Kits for O2Assured Claw Systems


              KIT NUMBER      DESCRIPTION                 QTY    WHERE USED              CONTENTS

            6-Month Basic Service Kit for 15 Hp O2Assured Claw Systems

                              KIT - Claw Vacuum Pump      1 per                          (1) High melting point
              4107 4001 25                                       15 Hp B-Side Bearings
                              Bearing Grease O2Assured    pump                           grease (400g tube)
           1-Year Basic Service Kit for O2Assured Claw Systems

                              KIT - Claw Vacuum Basic     1 per
              4107 4001 19                                       2 - 5.4 Hp              Filter element
                              Size A O2Assured            pump

                              KIT - Claw Vacuum Basic     1 per
              4107 4001 20                                       6.4 - 15 Hp             Filter element
                              Size B O2Assured            pump
           10,000 Hour Oil Change Kits for O2Assured Claw Systems

                                                                                         (1) Vent Valve
                              KIT - Claw Vacuum Oil       1 per
              4107 4001 21                                       2 - 5.4 Hp              (2 Pt) Synchronising
                              Change Size A O2Assured     pump
                                                                                         Gear Operating Fluid
                                                                                         (1) Vent Valve
                              KIT - Claw Vacuum Oil       1 per
              4107 4001 22                                       6.4 - 15 Hp             (2 Pt) Synchronising
                              Change Size B O2Assured     pump
                                                                                         Gear Operating Fluid
           Oil Analysis Kit for O2 Assured Claw Systems

                              KIT - Claw Vacuum Oil      1 per                           (1) Oil sample container
           4107 4001 27                                          All systems
                              Analysis O2 Assured        pump                            (1) Instruction form

           Replacement Parts
                                                         1 per
              4107 6541 85    Coupling Insert                    2 Hp - 7.5 Hp           (1) Coupling insert
                                                         pump

                                                         1 per   8.7 Hp - 10* Hp
              4107 6541 86    Coupling Insert                                            (1) Coupling insert
                                                         pump    (MM1322)
                                                         1 per
              4107 6541 87    Coupling Insert                    10* Hp (MM1402)         (1) Coupling insert
                                                         pump
                                                         1 per
              4107 6541 88    Coupling (Complete)                15 Hp                   (1) Complete coupling
                                                         pump


           Note: There are two models of 10 Hp Claw pumps: MM1322 and MM1402.  See pump nameplate to
           identify pump model.








           4107 9000 95.04                                  7-2