Page 291 - Womens Pavilion
P. 291
Oil-Less Claw Medical Vacuum
6.0 Maintenance
1 Pump Maintenance Components
5
1 Inlet Filter (Suction)
2 Discharge
2
3 Oil Drain Valve
4 Oil Sight Glass
6 5 Vent Valve / Oil Fill Port
3
4
6 Vacuum Relief Valve
2 - 10 Hp
Standard and O Assured
2 7 Back-Side Bearing Grease Fittings
1
5
6.1.1 Pumps
6
Maintain the pump regularly to achieve the
2
best operating results. Maintenance intervals
7 will depend on the pump’s use and ambient
conditions. Each pump in the LifeLine system has
an oil-lubricated gear box but there is no lubricant
in the pumping chamber.
3 4
Do not run pumps without oil in the gear box.
15 Hp Standard
6.1.2 Oil Type
1 a) Standard Claw - See Section 7.1 for listing of
5 oil change kits.
6 b) O Assured Claw - Use only BeaconMedæs
2
O Assured oil. Failure to use the proper
2
lubricant may void the warranty.
2
BeaconMedæs O Assured oil is inert in
2
the presence of oxygen that may be in your
vacuum system. See Section 7.1 for listing of
oil change kits.
7
4 Contact BeaconMedæs Tech Support for
3
additional information.
15 Hp O Assured
2
NOTE: If changing brands of oil, it is important to
Figure 6.1 Vacuum Pumps completely drain out all oil from the pump.
4107 9000 95.04 6-2