Oil-Less Claw Medical Vacuum


           5.0  Trouble Shooting



           Problem                     Possible Causes                   Solution

           Pump overheats              Cooling ducts blocked             Clean cooling ducts

                                       Cooling fan broken                Replace fan

                                       High ambient temperature          Ventilate or cool room
                                       Vacuum too high                   Adjust vacuum  settings or relief valve

                                       Inlet restricted                  Remove restriction

                                       Exhaust restricted                Remove restriction

                                       The gearcase may be low/ empty    If the oil level is found to be low,
                                       of oil, has contaminated oil or was   immediately shut off  the pump, drain
                                       fi lled with an incorrect type of oil.  the remaining oil from the gearcase and
                                                                         replenish with new fresh oil.  If an incorrect
                                                                         oil type or contamination is found, contact
                                                                         the factory for assistance.

                                                                         Refer to Section 5.1 for the correct oil type.

           Unit runs rough and cannot   Worn coupling                    Remove motor & inspect coupling element
           be rotated manually                                           Replace, if necessary

                                       Worn bearings                     Contact factory for assistance

           High vacuum level           Vacuum relief valves need adjusting  Contact factory for assistance

                                       Vacuum setting out of adjustment  Adjust vacuum setting on main display

           Excessive noise level       The coupling rubbers may be worn  See “Worn coupling” above

                                       Internal pump damage              Contact factory for assistance