Oil-Less Claw Medical Vacuum


           4.0  General Operation



           The Lag Alarm will activate if the vacuum transducer   4.3.3   Service Warnings
           signal circuit is lost or the transducer malfunctions.
           Press the alarm silence button on the main display   Service Due Alarm - Service intervals and type
           screen to silence the alarm (see Troubleshooting    of service are preprogrammed into the control
           Section 5). The vacuum level reading on the main    system. The background of the wrench icon on the
           display will read “ERR” when this alarm occurs.     main display screen toolbar will turn red when
           The vacuum system will continue to run off  the     one of these services are required. See Table 6.1
           backup vacuum switch.                               Maintenance Schedules.


           VSD Fault Alarm (VSD Systems only) - A VSD Fault
           Alarm occurs due to a malfunction of the VSD
           inverter.  Before resetting the alarm, investigate
           the cause of the alarm to make sure there is no
           damage to the VSD inverter.  See Section B.3 in
           Appendix B for VSD Inverter Trouble Shooting.  To
           reset, press the “Reset” button on the Alarm page
           of the main screen to enable the VSD to operate.

           VSD Off  Alarm (VSD Systems only) - A VSD Off
           Alarm occurs due to to the VSD inverter shutting
           down and becoming unavailable for operation.
           Before resetting the alarm, investigate the cause of
           the alarm to make sure there is no damage to the
           VSD inverter.  See Section B.3 in Appendix B for
           VSD Inverter Trouble Shooting and for instructions
           to restart the inverter.  To reset, press the “Reset”
           button on the Alarm page of the main screen to
           enable the VSD to operate.

           Ambient Temperature Alarm  - This alarm will
           activate when the temperature in the room
           exceeds the set point.  The touchxcreen will show
           an active alarm and record it in the event log.  The
           alarm remains latched until the alarm condition is
           reset by the operator.


           Cabinet Temperature Alarm (VSD Systems only)
           This alarm will activate when the temperature
           inside the control cabinet exceeds the set point.
           Check the side panel fi lters to make sure these are
           not clogged.  The touchscreen will show an active
           alarm and record it in the event log.  The alarm
           remains latched until the alarm condition is reset
           by the operator.