Oil-Less Claw Medical Vacuum


           3.0  Start Up



           Step 1:   Running Pump without VSD                  3.3  Initial Operation

           Check for correct direction of rotation of each     Start each pump by pressing “Automatic” on the
           pump by pressing the “Rotation” button on           touchscreen.  See Figure 3.5.
           the touchscreen display (found in the Service
           section of the Unit touchscreens) and observing
           rotation.  See Figure 3.4.  The Pump Mode for each
           compressor must be in the Off  Position for the
           Rotation to function.

           By observing the cooling fan of the motor, you
           can determine the rotation of the pump. Pump
           rotation should be counterclockwise when looking
           at the rear of the motor.  Directional arrows are
           located on each pump.

           If the pumps are rotating in the wrong direction,
           rotation can be reversed by switching any two
           main power leads to the panel.  Correct rotation
           should be confi rmed in the previous manner.         Figure 3.5 Unit Screen - Automatic Mode

           Step 2:   Running Pump with VSD                                        WARNING:

           Check for correct direction of rotation of each          Pumps that have reached operating
           pump by pressing the “Automatic” button on the           temperature may have a high surface
           touchscreen display and observing rotation.  See         temperature on the top of the exhaust
           Figure 3.5.                                              muffl  er.

           By observing the cooling fan of the motor, you                      DO NOT TOUCH!
           can determine the rotation of the pump. Pump
           rotation should be counterclockwise when looking
           at the rear of the motor.  Directional arrows are   Run the pump for two minutes in the correct
           located on each pump.                               rotation.  Stop the pump and check that the gear
                                                               box is at the correct oil level, as shown in the sight
           If the pumps are rotating in the wrong direction,   glass (See Figure 6.1). See Section 6.1.3 for the oil
           rotation can be reversed by switching any two        type and quantity.   DO NOT OPEN THE FILLER
           main power leads between the VSD and the motor       PORT WHILE THE PUMP IS RUNNING.
           starter.  Correct rotation should be confi rmed in
           the previous manner.                                 After testing each pump, if everything appears
                                                                normal, put each pump into the “Automatic”
                                                                mode and allow each pump to run until vacuum
                             WARNING:                           builds. Check for any leaks in the piping. Repair
                Do not allow the vacuum pump to                 leaks, if needed.
                           run backwards.