Oil-Less Claw Medical Vacuum


            3.0  Start Up



            3.1  Prestart-up                                     •  Check the intake piping for proper size and
                                                                    connection to the vacuum modules.
            The contractor should notify  BeaconMedæs
            two weeks prior to start-up date to schedule an      •  Check all piping system joints that might have
            appointment for an authorized technician to             come loose during shipment and installation
            review the installation prior to start-up.              to ensure they are tight.

            CAUTION:   Failure to install the unit properly      •  Check the air receiver, controls, and pumps for
            and have an authorized technician from                  damage.
            BeaconMedæs  start-up the system can void the
            manufacturer’s warranties.                           •  Check the drain valve on the air receiver.
                                                                 •  Check all valves for full open and full close
                               WARNING:                             travel. Ensure that the system’s valves are

             Prior to putting the LifeLine® Claw Medical            positioned for proper operation.  (Refer to
             Vacuum system into  use, the medical                   labeling on valve handles)
             facility must have a Certifi er perform all
             installation tests as specifi ed in NFPA 99.         •  Remove all packing material from the
             The medical facility is also responsible for           unit.
             ensuring that the Medical Vacuum meets
             the minimum requirements for Medical                •  Check the electrical connections to the control
             Vacuum as specifi ed in NFPA 99.                        cabinet.

                                                                 •  Verify electrical service. Before starting the
                                                                    system, check to see that voltage, amperage,
            Prestart-up and start-up procedures should be
            performed for a new installation or when major          and wire size are appropriate.
            maintenance has been performed.
                                                                 CAUTION: Electrical service must be as specifi ed
                                                                 or damage to equipment may occur.
                               WARNING:

             Have more than one person on hand during                              WARNING:
             prestart-up and start-up procedures to               To prevent electrical shock, ensure that
             ensure safety and to facilitate certain              ALL electrical power to the system is OFF,
             checks.
                                                                  including the disconnect switches and
                                                                  Automatic-Off -Manual touch screens on
                                                                  the control panel.  The facility’s supply
            The main power source to the control panel should
            be OFF for the duration of the visual inspection.     circuit breaker should also be locked out.

            Ensure that the equipment is installed on a solid    •  Open the electrical cabinet by loosening the
            level surface. Walk around the system to ensure that    fasteners on the front.
            there is enough clearance on all sides to perform
            operational checks/actions and maintenance. The      CAUTION: Vibration during shipment and
            temperature of the area containing the modules       installation can loosen electrical terminals, fuse
            should be approximately 70°F (21.1°C) with a         inserts, and mechanical connections. Tighten as
            minimum ambient temperature of 40°F (4.4°C)          necessary.
            and a maximum ambient temperature of 105°F
            (40°C).

                                                             3-1                                  4107 9000 95.04