Oil-Less Claw Medical Vacuum


            2.0  Installation


                Alarm         Data            Electrical         2.6  Intake Piping
                 Wires     Connection       Connection
                                                                 Before connecting any piping, the plastic thread
                                                                 protector installed in the connection port must be
                                                                 removed. We recommend that the main vacuum
                                                                 line to the receiver should not be reduced below
                                                                 that provided on the receiver. Long piping runs may
                                                                 need to be increased in size to minimize pressure
                                                                 drop. Improper line sizing may result in a loss of
                                                                 capacity. Ideally, piping should be constructed
                                                                 using long radius elbows and a minimum number
                                                                 of turns.

                                                                 All secondary lines should be taken from the top or
                                                                 side of the main line to prevent any accumulated
                                                                 moisture from draining towards the pumps. All
                                                                 lines should slope away from the pumps. Any low
                                                                 points in the piping should be equipped with pipe
                                                                 drains to remove accumulated moisture.

                                                                 All intake vacuum lines must be piped in
                                                                 accordance with NFPA 99. All pipe must be
                                                                 either seamless copper tubing or other corrosion-
                                                                 resistant metallic tubing, as detailed in NFPA 99.
            Figure 2.1 Electrical/Alarm/Data Openings
                                                                 2.7  Exhaust Piping
            Ensure that the emergency generation system
            electrical supply is consistent with the vacuum      The exhaust line must be piped outside of the
            system’s requirements.                               building in accordance with NFPA 99. To ensure
                                                                 that no restriction of airfl ow will occur, size the
            The electrical controls for the system were wired    piping according to Table 2.2. All pipe must be
            at the factory and were fully tested.                either seamless copper tubing or other corrosion-
                                                                 resistant metallic tubing as detailed in NFPA
            Three-phase power supplied from emergency            99. A fl exible connector must be installed on
            generator(s) must match that of the normal supply    each exhaust port of the vacuum pump before
            to allow for correct direction of the motor rotation   connecting to the main exhaust line leading
            at all times.                                        outdoors.  Additionally, a drip leg must be installed
                                                                 at each exhaust port connection to allow for the
            NOTE: It may be necessary to switch two of the       draining of any accumulated moisture (Refer to
            leads when performing start-up, if the pump          the installation schematics for more details).  The
            rotation is in the wrong direction.                  outside pipe must be turned down and screened
                                                                 to prevent contamination.










                                                             2-3                                  4107 9000 95.04