Page 317 - Womens Pavilion
P. 317
Oil-Less Claw Medical Vacuum
Appendix A: TotalAlert Embedded Control System
Maintenance Kits: The Maintenance Kits page
(Figure A.40) displays all service kits required
to perform the maintenance activities on the
Maintenance page. Included on the page are the
service item, the time interval for each time, and
the part number and description of the service kit
required to perform the maintenance function,
specifi c to the medical vacuum system installed.
Figure A.43 Device Information
Device Information: The Device Information page
(Figure A.43) displays information specifi c to the
TotalAlert Embedded device. The information
displayed includes an array of details pertaining
Figure A.41 Maintenance Log Page to the system, including model number, serial
number, contact information for service, and
Maintenance Log: The Maintenance Log page much more. The Device Information page includes
(Figure A.41) displays a listing of all maintenance specifi c information about the printed circuit
activity for both the system and the units. The boards found in the control cabinet (hardware
log displays the date/time of the event and a and software descriptions).
description of the event (maintenance due alert,
maintenance performed, interval reset). The
Maintenance Log page includes the option to
create a downloadable text fi le of the log.
Figure A.42 Network Devices
Network Devices: The Network Devices page
(Figure A.42) displays all TotalAlert and TotalAlert
Embedded devices on the facility’s network. The
page displays the device name, IP address, device
type, device serial number, and device location.
By clicking the IP address of a device, the user
moves to the website of that device.
4107 9000 95.04 A-20