Page 318 - Womens Pavilion
P. 318

Oil-Less Claw Medical Vacuum


            Appendix A:  TotalAlert Embedded Control System



                                                                 System Log: The System Log page (Figure A.45)
                                                                 displays a listing of all alarm and shutdown activity
                                                                 for both the system and the units.  The log displays
                                                                 the date/time of the event and a description of
                                                                 the event (alarm/shutdown, reset, test).  The
                                                                 System Log page includes the option to create a
                                                                 downloadable text fi le of the log.

                                                                 A.7.12   Available for Download

                                                                 In three of the viewing pages on the website, the
                                                                 user can download information for analysis or
                                                                 record keeping.  The Maintenance Log, System
                                                                 Trend, and System Log pages all contain a link for
                                                                 the user to click to create the downloadable fi le.

                                                                 1.  Click on one of the pages to display
                                                                    the information available for download
            Figure A.44 System Trend
                                                                    (Maintenance Log, System Trend, or System
            System Trend: The System Trend page (Figure             Log).
            A.44) displays multiple items on the system
            measured at specifi ed time intervals.  These items   2.  Click on the link to create a downloadable fi le,
            may include vacuum level, pump operation, and           located in the upper right above the displayed
            ambient temperature.  The time intervals may be         information.
            every 30 seconds, 2 minutes, 5 minutes, or 20
            minutes (see section A.7.6 System Trend Setup).      3.  Select to view the information on the web
            The System Trend page includes the option to            page or to save the fi le to the computer.
            create a downloadable spreadsheet fi le of the        The Maintenance Log and the System Log create
            events.  To clear the system trend data, go to the   text fi les that list all of the events in descending
            System Trend Setup page on the website (see          order from most recent to oldest.  See Figure A.46.
            Section A.7.6).
                                                                 The System Trend creates a CSV fi le (spreadsheet)
                                                                 that contains all of the trend information,
                                                                 decsending from most recent to oldest.  See Figure
                                                                 A.47.













            Figure A.45 System Log





                                                             A-21                                 4107 9000 95.04
   313   314   315   316   317   318   319   320   321   322   323