Oil-Less Claw Medical Vacuum


           Appendix A:  TotalAlert Embedded Control System



                                                               feature of the TotalAlert Embedded control
                                                               system.  By setting up the Electronic Notifi cation,
                                                               key personnel can receive notifi cations of all alarm/
                                                               shutdown alerts and/or all maintenance alerts
                                                               (e-mail, pager, phone text message).  The device
                                                               acts as an SMTP client.  An SMTP server is required
                                                               for electronic notifi cation to function.
           Figure A.33 Website Setup

           1.  Click Website Setup to access the Website
              Setup page.


           2.  Select a time interval from the pull-down list
              of options.

           3.  Click the Submit button.


           A.7.9   System Trend Setup





                                                               Figure A.35 Electronic Notifi cation Setup

                                                               1.  Click Electronic Notifi cation Setup to access
                                                                   the Electronic Notifi cation Setup page.

           Figure A.34 System Trend Setup
                                                               2.  Select Enable to enable the e-mail notifi cation
                                                                   tool.
           This System Trend Setup page (Figure A.34) is used
           to confi gure the trend log for the website and to
           allow the data to be cleared.  The System Trend     3.  Enter up to fi ve email addresses.
           screen contains a maximum of 720 events for each
           of the items recorded, so the actual duration of the   4.  For each address, select “System” for that
                                                                   person to receive all alarm/shutdown alerts.
           trend log changes with the time interval selected.      Select “Maintenance” for that person to
           The durations available to select are 6 hours, 24       receive all maintenance alerts.  An individual
           hours, 2.5 days, and 10 days.
                                                                   may receive both types of alerts.
           1.  Click System Trend Setup to access the System
              Trend Setup page.                                5.  Enter an email address in the “From” box as
                                                                   this will be the sender of the notifi cations.
           2.  Select a time interval from the pull-down list   6.  If desired, early warning notifi cations can be
              of options.
                                                                   set up for maintenance items.  By selecting
                                                                   these time intervals, a notifi cation will be sent
           3.  Click the Submit button.
                                                                   before the maintenance item is due, providing
           A.7.10   Electronic Notifi cation Setup                  time for planning the service activity.

           This Electronic Notifi cation Setup page (Figure     7.  Click the Submit button.
           A.35) is used to confi gure the Electronic Notifi cation

           4107 9000 95.04                                 A-18