Page 314 - Womens Pavilion
P. 314
Oil-Less Claw Medical Vacuum
Appendix A: TotalAlert Embedded Control System
A.7.7 Network Setup
This Network Setup page (Figure A.32) is used
to confi gure the facility network information and
e-mail server confi guration.
NOTE: Obtain the IP Address and DNS Name from
the facility’s Information Systems department.
Figure A.30 Setup Menu
A.7.6 Device Setup
This Device Setup page (Figure A.31) is used to
confi gure the medical vacuum system name, Figure A.32 Network Setup
location, facility name, and contact information.
1. Click Network Setup to access the Network
Setup page.
2. Select Enable or Disable for the DHCP Mode.
3. Enter the IP Address for the DNS server
confi guration.
Figure A.31 Device Setup
4. Enter the IP Address and DNS name for the
1. Click Device Setup to access the Device Setup e-mail SMTP server confi guration.
page.
5. Click the Submit button.
2. Enter the new device name.
A.7.8 Website Setup
3. Enter the location
This Website Setup page (Figure A.33) is used to
4. Enter the facility description. confi gure the automatic refresh interval for the
website. By selecting an auto-refresh interval, the
5. Enter the contact information. web page will be refreshed automatically, keeping
all information current during a viewing period.
6. Click the Submit button.
A-17 4107 9000 95.04