Page 24 - Zavation MIS Retractor System - Hensler Surgical 2019
P. 24
Retractor Kit Rev: B
Instructions for cleaning, sterilization and maintenance
2.3. Sterilization
Please refer to chapter 2.1.5.Table clamp base special handling care to free up
the washer of the table clamp base before sterilization.
Regarding the extension cords, the light cables and the table clamp, refer to the
corresponding Instruction For Use (IFU).
Sterilization should be done following the following procedure:
steam sterilization, pre-vacuum cycle, 121°C (250F), 30min
DO NOT STERILIZE THE SINGLE-USE PRODUCTS.
2.4. Storage
The devices have to be stored in dry conditions: area shall be ventilated, safe from dust,
humidity, insects and others pests, any other potentials contamination sources. No
extreme temperature and humidity rate are allowed.
The user have to prevent any mix up, damage, deterioration, contamination, or another
adverse effects to the products during the handling and the storage.
Regarding the probe, the light mats, the extension cords, the light cables and the table
clamp, refer to the corresponding Instruction For Use (IFU). Be aware that the storage
dwell time of sterile products is limited by the sterilization expiry date.
24