Retractor Kit Rev: B
Instructions for cleaning, sterilization and maintenance
2.2.3. Automatic Cleaning
Completely submerge the device in an enzyme solution (Aniosyme DD1, 0.5%) and allow
to soak for 5 minutes at room temperature (20-25°C) with ultrasonic vibrations.
Use a soft bristled, nylon brush to gently scrub the device for 2 minutes at room
temperature (20-25°C) and until all visible soil has been removed. Do not use metallic
brush.
Particular attention must be given to crevices, lumens, mated surfaces,
connectors and other hard-to-clean areas.
Thoroughly flush lumens, holes and other hard-to-clean areas with the enzyme solution
(Aniosyme DD1, 0.5%).
Rinse device in purified water for at least 1 minute. Thoroughly flush lumens, holes and
other hard-to-clean areas.
Place the devices in a suitable washer/disinfector basket and process through a standard
washer/disinfector cleaning cycle.
Visually inspect the device for remaining soil and repeat the steps above if necessary.
Example of Vario TD cleaning cycle:
HAMO ECOLINE LM-25
Cycle Vario TD liquide
NEODISHER FLA
Pre-cleaning <45°C 2 minutes
Cleaning 55°C 5 minutes
Neutralization 2 minutes
Rinsing Tap water
2 minutes cold water
Thermal 90°C 5 minutes
Disinfection Demineralized water
Drying 22 minutes
DO NOT CLEAN THE SINGLE-USE PRODUCTS.