Retractor Kit Rev: B
Instructions for cleaning, sterilization and maintenance
PAGE TO BE REMOVED FROM FINAL CUSTOMER INSTRUCTIONS FOR USE
- With the retractor body in its opened position (see figure 28), remove the screws on
the bottom side of the retractor body: the central screw will release the central arm (see
figure 29), left and right screws release left and right arms (see figure 30).
BEWARE NOT TO LOSE ANY PART (SCREWS, SPRINGS, ETC)
THIS DISASSEMBLING IS NOT RECOMMENDED
FOR MEDICAL HEALTHCARE FACILITIES
Figure 30
Figure 28
Figure 29