Page 20 - Zavation MIS Retractor System - Hensler Surgical 2019
P. 20
Retractor Kit Rev: B
Instructions for cleaning, sterilization and maintenance
2.1.2. Shim/broach holder disassembling
Disassemble the shim/broach holder for cleaning: take off the pin (see figure 31), then
pull the firing pin and hold up the rod (see figure 32).
Figure 31
Figure 32
th
2.1.3. 4 blade support special handling care
The support must be completely opened (maximum wingspan) to maximize the access
for cleaning. Screw must be totally loosened.
Figure 33