F.  Rotors

                       The function of the Rotor is to contain and spin-sweep the bone chips over the surface of the
                       blade. Revolving Rotor generates centrifugal forces that push the bone chips against the surface
                       of the blade allowing the cutting teeth of the blade cut portions of the bone chips. Each Rotor is

                       designed to function in conjunction with specific group of blades.



                      S16 Rotor

                      For S-Blades
                                                                  Autoclave
                      -  The S16 Rotor specifically designed to
                        match the size-and-cut S-blades that are
                        for milling cancellous or cortical-

                        cancellous (not pure cortical) bone.
                      -  S16 Rotor is capable of accepting 3
                        bone chips as large as 16mm in

                        dimensions (12cc total) per loading.
                                                                  S16 Rotor






                      M16 Rotor                                   Autoclave


                      For the self-clearing M-Blades

                      The M16 Rotor specifically designed to
                      match the cut-and-trim blades that are for
                      milling cortical or cortical-cancellous
                      (not pure cancellous) bone.

                      - M16 Rotor is capable of accepting 3

                        bone chips as large as 16mm in
                        dimensions (12cc total) per loading.      M16 Rotor