Page 9 - HR500 User Guide - 2016-USA Version Hensler_Neat
P. 9

8.  HR500 Bone Mill Components

                  A.  Sterile Motor Frame, Autoclavable
                       The basic function of the Motor Frame, shows in Figure 2, is to provide housing for the non-

                       sterile battery-power-pack and safely provide the rotational power to the bone milling chamber.
                       Figure 3 shows the clamshell design of the Motor frame comprising the Motor frame and the
                       Base. The Base includes an “O” ring seal arrangement that isolates the non-sterile battery-power-

                       pack from the operating room sterile environment, when the clamshell is closed and locked. The
                       Front cam is designed to close-lock the Motor frame to the Base.



                       Motor Frame                                Clamshell Design




                                                   ----------Motor frame








                                                                                 --------Battery-power-pack

                                                    -Contacts----
                                                                                                -----“O” Ring

                                                                                                  --Front cam

                                     Hinge----------------                                     ------Base
                       Figure 3













               8
   4   5   6   7   8   9   10   11   12   13   14