Normandy VBR System Surgical Technique Guide ST-017, rev 1
             Step 10 – Lock Screw Insertion


             Press the self-retaining distal tip of the lock
             screw inserter instrument into the lock screw.


             Position the implant inserter instrument,
             without the attachment shaft, into the implant
             attachment slots.  Using the implant inserter
             instrument as an alignment guide, insert the
             lock screw through the cannula of the implant
             inserter until the lock screw rests against the
             implant.  Twist the lock screw inserter
             instrument clockwise until the lock screw is
             hand tight.  Remove the lock screw inserter
             shaft.





             Attach the lock screw torque shaft to the
             torque limiting handle.  Insert the torque shaft
             through the implant inserter cannula and rotate
             until the torx tip of the torque shaft aligns to
             the torx drive of the lock screw.  The torque
             shaft will advance into the lock screw when
             properly aligned.  Using the implant inserter
             instrument as an anti-torque, hold the inserter
             handle (not the expander knob) and rotate the
             torque limiting handle clockwise until the
             handle clicks indicating the required torque
             has been placed on the lock screw.








                                                                         Inserter Handle – hold
                                                                         here for anti-torque