Normandy VBR System Surgical Technique Guide ST-017, rev 1
Step 7 – Implant Attachment
Align the prongs on the distal end of the
inserter to the slots on the implant. Turn the
expander know to align the teeth of the
inserter instrument to the teeth of the implant.
When the teeth are aligned the tips of the
prongs will be fully engaged into the slots of
the implant as shown in the image below.
After verifying the teeth are aligned, insert the
inserter attachment shaft through the center
cannula and thread the shaft into the implant
until the head of the shaft sits against the
expander knob.
Gap indicates the No gap indicates
gear teeth are not the gear teeth are
aligned
aligned and the
inserter is fully
seated against the
implant