Page 55 - Feed Lab
P. 55

NIR Analysis Procedure                        Suguna Management System      Ver 1.0 / SOP / FQL / P1 - 10  Page  2 of 3


                     of  major  and  minor  sample               PROCEDURE FOR CHECK CELL
                     components;                                 ANALYSIS BY NIR SYSTEM

                 b)  Seasonal,  geographic  and  genetic
                                                                               From the Opening menu
                     effects on   forages, feed raw material
                     and cereals;
                                                                                  Click on Analysis
                 c)  Processing techniques and conditions;

                 d)  Storage conditions; and
                                                                           Highlight Scan Samples and Click
                 e)  Sample and instrument temperature.


           FCalibrations must be provided with statistical                           Press Enter
              information  on  its  performance  such  as
              number  and  type  of  samples,  correlation,
              standard error of calibration and prediction and                 Highlight on Check Cel
              in a format that allows for the identification of
              outliers.
                                                                                Highlight on Check Cel
           Reference Analyses and NIR measurements
           FInternationally accepted reference methods for                          Press Enter
              determination  of  moisture,  fat,  protein  and
              other constituents and parameters should be
              used.                                                         Highlight on Analyze Check Cell

           FThe  reference  method  used  for  calibration
              should  be  in  statistical  control,  i.e.  for  any                 Press Enter
              sample,  the  variability  should  consist  of
              random variations of a reproducible system. It
              is  essential  to  know  the  precision  of  the                The Check Cell Scanning
                                                                            Starts and after a few Minutes
              reference method.                                                  it will be completed




                                                                           Highlight on Standard Check Cell

           Note :
                                                                                    Press Enter
              The Check cell should be scanned daily and before
           F
              Scanning  the  Equipment  it  should  be  kept  in
              Analysis mode at least for 30 mins.
                                                                               Result will be displayed
           FOn  the  instrument  check  the  Check  cell
              performance after analysis unknown sample.
                                                                                    Press Enter
           FMonthly basis analyse the Diagnostics. And check
              the Performance of Accuracy & Repeatability of the
              NIR Instrument.





          Prepared by :                                                             Approved by :
                                                             45                                    Director - Technical