Page 161 - 25270_1
P. 161

11




           NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.

           NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation
 Guidance and manufacturer’s declaration – electromagnetic immunity for not life   is affected by absorption and reflection from structures, objects and people.
 support equipment
           a
 The Product is intended for use in the electromagnetic environment specified below.   Fixed strengths from fixed transmitters, such as base stations for radio (cellular/
 The customer or the user of the Product should assure that it is used in such an   cordless) telephones and land mobile radios, amateur radio, AM and FM radio
 environment.
           broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
 Immunity   IEC 60601   Compliance   Electromagnetic environment - guidance  access the electromagnetic environment due to fixed RF transmitters, and
 tests  Test level  level  electromagnetic site survey should be considered. If the measured field strength in the
 Portable and mobile RF communications   location in which the Product is used exceeds the applicable RF compliance level
 equipment should be used no closer to   above, the Product should b observed to verify normal operation. If abnormal
 any part of the Product, including cables,   performance is observed, additional measures may be necessary, such as reorienting
 than the recommended separation   or relocating the Product.
 distance calculated from the equation
 applicable to the frequency of the   b
 transmitter.  Over the frequency range 150 kHz to 80 MHz, field strengths should be less than
           3 V/m.
 Recommended separation distance:
 d = 0.35 √P    80 MHz to 800 MHz
 d = 0.7 √P    800 MHz to 2.5 GHz
 Where P is the maximum output power
 rating in the transmitter in watts (W)
 Radiated RF   3 V/m  3 V/m  according to the transmitter
 IEC  80 MHz to   80 MHz to   manufacturer and d is the recommended
 61000-4-3  2.5 GHz  2.5 GHz
 separation distance in metres (m).

 Field strengths from fixed RF
 transmitters, as determined by an
 a
 electromagnetic site survey  , should be
 less than the compliance level in each
 b
 frequency range  .
 Interference may occur in the vicinity of
 equipment marked with the following
 symbol:











 162/172  B EU PIIL DFU PRT / Rev.07 / 02-2018 (Old ZF 190292.X)  B EU PIIL DFU PRT / Rev.07 / 02-2018 (Old ZF 190292.X)  163/172

























































 10


























                                                                                        # 25270-EYAL - 25270_pnim_eyal.1.1 | 6 - A | 18-05-15 | 13:28:56 | SR:-- | Magenta  25270-EYAL - 25270_pnim_eyal.1.1 | 6 - A | 18-05-15 | 13:28:56 | SR:-- | Yellow  #25270-EYAL - 25270_pnim_eyal.1.1 | 6 - A | 18-05-15 | 13:28:56 | SR:-- | Black