NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation
Guidance and manufacturer’s declaration – electromagnetic immunity for not life is affected by absorption and reflection from structures, objects and people.
support equipment
a
The Product is intended for use in the electromagnetic environment specified below. Fixed strengths from fixed transmitters, such as base stations for radio (cellular/
The customer or the user of the Product should assure that it is used in such an cordless) telephones and land mobile radios, amateur radio, AM and FM radio
environment.
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
Immunity IEC 60601 Compliance Electromagnetic environment - guidance access the electromagnetic environment due to fixed RF transmitters, and
tests Test level level electromagnetic site survey should be considered. If the measured field strength in the
Portable and mobile RF communications location in which the Product is used exceeds the applicable RF compliance level
equipment should be used no closer to above, the Product should b observed to verify normal operation. If abnormal
any part of the Product, including cables, performance is observed, additional measures may be necessary, such as reorienting
than the recommended separation or relocating the Product.
distance calculated from the equation
applicable to the frequency of the b
transmitter. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than
3 V/m.
Recommended separation distance:
d = 0.35 √P 80 MHz to 800 MHz
d = 0.7 √P 800 MHz to 2.5 GHz
Where P is the maximum output power
rating in the transmitter in watts (W)
Radiated RF 3 V/m 3 V/m according to the transmitter
IEC 80 MHz to 80 MHz to manufacturer and d is the recommended
61000-4-3 2.5 GHz 2.5 GHz
separation distance in metres (m).
Field strengths from fixed RF
transmitters, as determined by an
a
electromagnetic site survey , should be
less than the compliance level in each
b
frequency range .
Interference may occur in the vicinity of
equipment marked with the following
symbol:
162/172 B EU PIIL DFU PRT / Rev.07 / 02-2018 (Old ZF 190292.X) B EU PIIL DFU PRT / Rev.07 / 02-2018 (Old ZF 190292.X) 163/172