Page 366 - The Veterinary Laboratory and Field Manual 3rd Edition
P. 366

Clinical chemistry  335


                •  The required graph is prepared by plotting the   colour range absorption recommended in the
                  absorbance (OD) values obtained against the   protocol selected for a particular test.
                  known concentration of the standard reagent.  •  Equipment should be calibrated using a con-
                                                           trol or standard reagent and a ‘blank’ sample
                If there is a linear relationship between the   prior to running the test.
                concentration of a test reagent and colour   •  As part of the quality control process, every
                absorption (OD) then the concentration of the   batch of samples for a biochemistry test must
                test substance is directly proportional to the   be run along with standard controls, that is, a
                colour intensity of the solution. However, in   known positive and a known negative control.
                most cases the relationship will be exponential   In the case of quantitative tests, a dilution
                (Figure 7.6) and there will only be a specific area   series of the known standard reagent should
                on the linear part of the ‘standard curve’ where   also be run.
                accurate readings can be taken. In these cases,
                very low or high values may not be valid and
                concentrated solutions or samples should be  quality assessment
                diluted and re-tested.
                                                         It is standard procedure to periodically send
                                                         duplicate serum/plasma samples to other lab-
                Summary                                  oratories for biochemical tests to check that
                                                         laboratory results are reliable. It is also impor-
                •  The colorimetry equipment used must be set   tant to run an internal quality control scheme by
                  at the correct wavelength to read the specific   keeping a record of all test results and preparing





















                Figure 7.7  Plotting values for quality control (QC). The method of recording QC results illustrated above
                allows detection of abnormal results which occur due to laboratory error. The QC samples are samples
                which contain a known amount of test substance and are run along with each batch of test samples.
                The example above shows that samples tested in months 4–5 gave unacceptably low values but that
                the problem appears to have been sorted out by month 6. It is important to detect potential problems as
                early as possible and therefore monitoring of QC samples should be done on a daily basis in a very busy
                laboratory. The acceptable and ‘normal’ values for each set of QC samples should be determined for each
                test performed in the laboratory and a portion of the QC sample used should be sent to another diagnostic
                laboratory to make sure that values are correct.







       Vet Lab.indb   335                                                                  26/03/2019   10:26
   361   362   363   364   365   366   367   368   369   370   371