Page 89 - The Veterinary Laboratory and Field Manual 3rd Edition
P. 89
58 Willy Schauwers
For solar-powered cooling equipment you
should also record the state of charge level of
the batteries (as a percentage).
Temperature registration quality
assurance
See ‘SOP-LABORATORY-EQUIP-05 Temperature
registration’.
tHErMoMEtErS and rEcordInG EquIPMEnt
Many different types of thermometers exist:
• the common thermometer (always use ther-
mometers without mercury, because the
mercury vapours – released when a thermom-
eter is broken – are hazardous)
• certified thermometers
• minimum–maximum thermometers: ther-
mometer with gliders (can be reset by using
a magnet) (Figure 2.17)
• digital thermometer with internal and exter-
nal temperature measuring probe (Figure
Figure 2.17 A maximum–minimum thermometer. 2.18)
The scales are Fahrenheit on the inside of the U • digital thermometer with data logger:
and Celsius on the outside. The current tempera-
ture is 23 degrees Celsius, the maximum recorded • data logger with internal temperature
is 25°C, and the minimum is 15°C. Photo: Lumos3 probe (Figure 2.19) (the device has to be
– Wikipedia. put in the equipment you want to mea-
sure the temperature from, for instance a
refrigerator)
• data logger with multiple temperature
probes (connected to the data logger), one
data logger can monitor the temperature
of one or several (nearby) devices (Figure
2.20)
• a wireless temperature monitoring device
which registers the temperature continuously
and is able to send an SMS or email when
the equipment temperature exceeds certain
Figure 2.18 Digital thermometer min/max with limits
external probe. Photo: Willy Schauwers, Provincial • chemical temperature monitoring devices,
Institute for Hygiene, Antwerp, Belgium. which register temperature to detect rising
Vet Lab.indb 58 26/03/2019 10:25