Page 93 - The Veterinary Laboratory and Field Manual 3rd Edition
P. 93

62  Willy Schauwers

              be allowed to come down slowly (to atmo-  •  Close the wing nuts properly (normal manual
              spheric  pressure  and  temperature  around   tightness is adequate).
              80°C) to prevent liquids inside the autoclave   •  Allow the autoclave to cool to 80°C or less
              from boiling and spilling. It is important to   and check that the pressure gauge is reading
              make sure that the autoclave is not still under   zero before opening the lid.
              residual pressure before opening it.   •  The use of alkalis (for example, sodium bicar-
                                                       bonate) for cleaning may cause corrosion
            Articles for autoclaving should be cleaned   of the pot and lid and should therefore be
            (before autoclaving) of any residual organic   avoided. Use normal household detergent to
            material.  Timing of  the autoclaving  process   clean the interior of the autoclave.
            should start when the sterilization pressure   •  An annual safety check will normally be
            (15 psi) is reached.                       sufficient  unless  the  autoclave is  in  very
              A SOP of a portable Dixons autoclave  can   regular use. See ‘FORM-SOP-LABORATORY-
                                              4
            be found in ‘SOP-LABORATORY-EQUIP-04       EQUIP-04–01 Autoclave Dixons Inspection’.
            Autoclave Dixons ST19T and ST3028’.      •  Electric autoclaves consume easily 2000 W,
                                                       therefore the electric wiring system should
                                                       be of sufficient quality to avoid electric
            Standard care and maintenance of a
            simple bench top autoclave                 failure.

            •  Fill the chamber with distilled water.  Quality control
            •  Place the load properly prepared for steril-
              ization in the inner liner, into a wire mesh   •  Check the silicone gasket for distortion or
              basket or a bottle rack.                 wear and tear.






















            Figure 2.23  Autoclave tape   Figure 2.24  Autoclave tape   Figure 2.25  Autoclave tape
            before sterilization. Photo: Dr Ziay  after sterilization. Photo: Dr Ziay   after sterilization (prominent
            Ghulam, Dr Wahidullah Bahaer,   Ghulam, Dr Wahidullah Bahaer,   black lines). Photo: Dr Ziay
            Central Veterinary Diagnostic   Central Veterinary Diagnostic   Ghulam, Dr Wahidullah Bahaer,
            and Research Laboratory, Kabul,   and Research Laboratory, Kabul,   Central Veterinary Diagnostic
            Afghanistan.               Afghanistan.               and Research Laboratory, Kabul,
                                                                  Afghanistan.