The selection, use, maintenance and quality control of laboratory equipment and supplies   61


                  To ensure that an autoclave is working   •  Heating up of the water and air removal.
                efficiently it is necessary to check that the tem-  The water is heated up to its boiling point
                perature/pressure readings are correct. The   of 100°C. During this heating up the air is
                usual method to verify this is to apply special   being removed from the chamber. In order
                autoclave tape to objects for sterilization. This   to improve the removal of air, the water is
                strip of self-adhesive ‘autoclave tape’ is impreg-  allowed to boil some additional time after
                nated with a chemical that changes colour if the   the water has reached 100°C. The air in the
                correct temperature/time has been employed.     chamber will be flushed out with the steam.
                A large range of chemical sterilization products   Care should be taken to expel all air from
                is available, such as Bowie and Dick test sheets,   the autoclave before it is put under pressure
                helix test.                                otherwise pockets of air will hinder penetra-
                                                           tion of steam into the articles to be sterilized.
                                                         •  Building up pressure. Temperature increases
                The basic steam sterilization process
                                                           to sterilization temperature. By closing a
                Air is about twice as heavy as pure steam of the   valve, the vessel is sealed and pressure and
                same temperature. Therefore, empty containers   temperature increase to the required level.
                should be put upside down, the heavier air will   •  Sterilization time (holding time). During this
                sink down and will be replaced by the lighter   time, the temperature/pressure is held at the
                steam (downward displacement). In order to let   required level for sterilization.
                the steam reach all parts, leave sufficient space   •  Reducing the pressure to atmospheric by
                between the goods.                         releasing the steam. By opening a valve the
                  A full sterilization cycle based on the removal   steam can escape and the pressure goes down.
                of air by downward displacement will go through   •  Cooling down of the load. After sterilization
                the following phases (Figure 2.22).        the pressure and the temperature should































                Figure 2.22  Sterilization cycle in graphic format.