Contaminations of the environment from the filter elements have
                                                to be taken care of immediately. An extraction and filtering device
                                                with at least the same filter class and an appropriate aspiration
                                                has to be used.
                                                 Attention: For your own safety, you should wear respiratory
                                                equipment  and  gloves  minimum  during  any  maintenance
                                                and upkeep work.
                                                Attention: Disconnect the unit from the mains supply before
                                                starting service and maintenance work!
                                                After  a  certain  period  of  time  the  activated  carbon  in  the
                                                adsorption  filter  cassettes  will  be  saturated  depending  on  the
                                                extraction  volume  of  gases  and  fumes.  An  indicator  for  the
                                                saturation  of  the  filters  can  be  the  characteristic  smell  of  the
                                                exhaust  gases  in  the  clean  gas  stream.  To  prevent  this  from

                                                happening  the  filter  cassettes  have  to  be  replaced  periodically
                                                after a set time interval before they are completely saturated.
                                                For the replacement of filter cassette proceed after the following
                                                instructions:
                                                1.  Open  the front  door  at  the  recessed grip.  At  LEF  300  open
                                                   both front doors.






















                                                2.  Unfasten the two screw handles on the left side of the filter.