Handling and Storage This extracting and filtering unit is designed for upright transport.
For easy transportation the device is equipped with castors.
Warning:
Keep the following conditions for handling and storage:
Temperature: -25 °C to +55 °C
Short-time value (max. 24 hrs.): up to +70 °C
Air humidity: not condensing
Installation
Mechanical Installation
Prior to commence installation please check this unit for any
possible transport damages.
The extraction unit has to be installed on level, horizontal and solid
ground.
After moving the device to its destined place the levelling feet must
be screwed down to lift the castors off the ground. Unscrew the
levelling feet on the upper nut until the device is evened out. Now
screw the nut in the upper direction and fasten it at the machine
housing.
To other machines, devices and parts of the building minimum
distances have to be met to ensure save maintenance works.
Following minimum distances have to be met:
Frontside: + 0,8 m
Left and right side: + 0,1 m
Electrical Connection
The extraction unit is provided with complete electric equipment.
The corresponding connection values can be found in the chapter
“Technical Specifications” and in the circuit diagrams.
Please note:
The interfaces are only intended for installation and dismantling of
the system. They may not be disconnected under load.
ACD TD - issue 09/2017 page 15