              4.6.1.  Alert Thresholds

                        GuardII+ SM Alerts can be set manually, or they can be automatically set by the monitor;
                     this is called the Alert Auto-Learn mode.

                                     NOTE:
                                     Do not use the Auto-Learn function unless the machine being monitored
                                     is running.

                        Auto Learn Mode is enabled in IAM during configuration (refer to Section 6.3 for details).
                     When Auto-Learn mode is enabled, the following occurs:
                        •  The monitor is put online.

                        •  Measurements are taken and archived as normal.
                        •  For a period of 24 hours, the measurement data is analyzed, and Alert thresholds are
                            set for each input based on the data collected.
                        •  After the Auto-Learn period, normal operation resumes.

                        After the 24-hour Auto-Learn period, the Alert levels are set as follows:
                        •  Each voltage and current input has a maximum Alert threshold set to 3x the average
                            data collected over the Auto-Learn period.
                        •  Each current input has a minimum Alert threshold set to 10% of the average data
                            collected over the Auto-Learn period.
                                     NOTE:
                                     The  maximum  Alert  thresholds  will  be  constrained  by  the  maximum
                                     sensitivity of the input.  For example, if average data collected over the
                                     Auto-Learn period for the TV input was 60 V then the calculated maximum
                                     Alert threshold would be 180 V.  However, if the TV input sensitivity is only
                                     80 V, then maximum Alert threshold is set to 80 V.

                                     NOTE:
                                     If the Auto-Learn process is interrupted by a power cycle, the monitor will
                                     start a new Auto-Learn when it is turned back on.

                                     NOTE:
                                     Once the Auto-Learn period is complete, use the ‘Synchronize’ function in
                                     IAM to import the new Alert thresholds into the database.


              4.6.2.  Alert Relay

                        GuardII+ is equipped with a solid-state Alert Relay that can be programmed with one of
                     five behavior options using IAM:

                        •  Disabled – the Alert Relay is not used.
                        •  Normally Open, Latching – if there is no Alert condition, the relay contacts are open.
                            When an Alert is detected the relay will close and stay closed until reset.