UV-visible Spectrophotometry
1- Using standard absorptivity value
A=εbC
A = A(1%,1 cm) b C
• This method is used for official stable compounds where ε and
A(1%,1 cm) are given by the pharmacopoeia.
• Advantage:
No need to prepare a standard solution of the reference substance.
This is important especially if the reference substance is not
commercially available or expensive.
• Disadvantage:
The measured absorbance of a given solution will usually vary
somewhat from one instrument to another, so the instrument should
be calibrated.
63
2- Using single point standardization
Procedure: measure the absorbance at λmax for the sample solution
and for a standard solution of the reference substance. The standard
and sample solutions are prepared in the same solvent. The sample
concentration can be calculated by the following equation:
Atest = a b ctest Astandard = a b cstandard
─A──tes─t = ───A──st─andard
ctest cstandard
ctest = ───A─t─es─t ─x─c─s─tandard
Astandard
64