Page 34 - Instrumental Analysis - Pharm D Clinical- 07-PA403
P. 34

UV-visible Spectrophotometry

1- Using standard absorptivity value

                                 A=εbC

                          A = A(1%,1 cm) b C

• This method is used for official stable compounds where ε and
    A(1%,1 cm) are given by the pharmacopoeia.

• Advantage:
    No need to prepare a standard solution of the reference substance.
    This is important especially if the reference substance is not
    commercially available or expensive.

• Disadvantage:
    The measured absorbance of a given solution will usually vary
    somewhat from one instrument to another, so the instrument should
    be calibrated.

                                                                                                    63

2- Using single point standardization

Procedure: measure the absorbance at λmax for the sample solution
and for a standard solution of the reference substance. The standard
and sample solutions are prepared in the same solvent. The sample
concentration can be calculated by the following equation:

Atest = a b ctest             Astandard = a b cstandard

─A──tes─t = ───A──st─andard
       ctest                  cstandard

ctest  =                      ───A─t─es─t ─x─c─s─tandard
                                     Astandard

                                                                               64